Total Organic Carbon Analyzer BK-TOC5000RN
Brand BIOBASE
Product origin CHINA
Delivery time Within 7 days
Supply capacity Factory Direct Supply
Model BK-TOC5000RN BK-TOC5000RD
Introduction:
The Total Organic Carbon (TOC) Analyzer is a high-precision instrument designed to measure the amount of organic carbon present in water or liquid samples. It is essential for ensuring water purity in industries such as pharmaceuticals, environmental monitoring, and chemical production. The TOC Analyzer detects organic contamination by measuring carbon levels, providing accurate and reliable results for compliance with regulatory standards and quality control protocols.
Download
Features:
1. Auto sample dilution, auto acid-adding and auto gas purging(BK-TOC5000RD).
2. 680℃ catalytic oxidation technology with platinum catalyst, especially for seawater test.
3. NDIR detector, high sensitivity and stability.
4. Optional autosampler.
5. Multi-functional PC software.
6. Precise gas flow control technology improves accuracy.
7. PID temperature controlling technology, ensure higher accuracy.
8. Unique three-stage dehydration technology improves drying efficiency.
9. Personalized standard curve management provides great convenience for users.
10. Consumables management reminds users to replace consumables in time.
11. Modular design, simplifying device operation and maintenance.
Technical Parameters:
Model | BK-TOC5000RN | BK-TOC5000RD |
Measurement Method | High temperature catalytic combustion | |
Detector | NDIR | |
Analysis Parameter | TC, TIC, TOC, NPOC | |
Control Mode | PC software controlled | |
Gas Requirement | Oxygen, purity ≥99.995% | |
Sample Type | Liquid sample(AS-W20 is optional.) | |
Measurement Range | 0~1000mg/L, can extend to 0~100,000mg/L(Manually dilution) | 0~1000mg/l, can extend to 0~100,000mg/L(Automatically dilution) |
Limit of Detection | TC: 50μg/L IC: 20μg/L | |
Measuring Time | TC: around 4min IC: around 3min | |
Max. Permissible Error | TOC: ±5% IC: ±4% | |
Repeatability | ≤3% | |
Injection Volume | TC: 100~500μl IC: 100~2000μl | |
Optional Accessories | Autosampler for liquid sample or solid sample Enhanced Halogen Removal Device Enhanced Halogen Removal Device-Metal Wire Mesh | |
Power Supply | AC220V±10%, 50/60Hz(Standard); 110V±10%, 50/60Hz(Optional) | |
Consumption | 700W | |
Package Size(W*D*H) | Main Machine: 780*710*610mm Accessories: 710*230*600mm | |