MicroPette Plus Autoclavable Pipette
Brand BIOBASE
Product origin CHINA
Delivery time Within 7 days
Supply capacity Factory Direct Supply
Introduction:
The MicroPette Plus Autoclavable Pipette are precision-engineered manual pipettes designed to deliver accurate and consistent liquid measurements for a variety of laboratory applications.
Download
Fully Autoclavable:
The MicroPette plus pipettes can be fully autoclaved at 121℃, 1 atm for less than 20 mins. The pipettes can be autoclavedwithout special preparations. After autoclaving the pipette must be cooled down and dried over 12 hours.
Features for MicroPette plus:
1. Fully autoclavable.
2. Light weight, ergonomic, low force design.
3. Digital display clearly reads volume setting.
4. The pipettes cover volume range of 0.1μl to 10ml.
5. Easy to calibrate and maintain.
6. Design helps avoid repetitive strain injuries.
7. Manufactured from innovative materials.
Features for multi-channel MicroPette plus:
1. 8 and 12 channel pipettes are available for standard 96-well plate.
2. Individual piston and tip cone assemblies allow easy repair and maintenance.
3. Compound material tip cone design allows visual seal verification.
4. Universal pipette tips can be used.
5. Dispensing head rotates for optimum pipetting convenience.
Note: It is recommended to check the performance of the pipette after each autoclaving. lt is also recommended to grease the piston and seal of the pipette after each 10 autoclaving.
Technical Parameters:
TopPette Mechanical Pipettes /MicroPette Plus Autoclavable Pipettes(Adjustable and Fixed Volume)
Single-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume(μl) | Maximume Permissible Systematic Error | Maximume Permissible Random Error | ||
0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.0625μl | 2.00% | 0.05μl |
1.25 | 3.00% | 0.0375μl | 3.00% | 0.0375μl | ||
0.25 | 12.00% | 0.03μl | 6.00% | 0.015μl | ||
0.5-10μl | 0.1μl | 10 | 1.00% | 0.1μl | 0.80% | 0.08μl |
5 | 1.50% | 0.075μl | 1.50% | 0.075μl | ||
1 | 2.50% | 0.025μl | 1.50% | 0.015μl | ||
2-20μl | 0.5μl | 20 | 0.90% | 0.18μl | 0.40% | 0.08μl |
10 | 1.20% | 0.12μl | 1.00% | 0.1μl | ||
2 | 3.00% | 0.06μl | 2.00% | 0.04μl | ||
5-50μl | 0.5μl | 50 | 0.60% | 0.3μl | 0.30% | 0.15μl |
25 | 0.90% | 0.225μl | 0.60% | 0.15μl | ||
5 | 2.00% | 0.1μl | 2.00% | 0.1μl | ||
10-100μl | 1μl | 100 | 0.80% | 0.8μl | 0.15% | 0.15μl |
50 | 1.00% | 0.5μl | 0.40% | 0.2μl | ||
10 | 3.00% | 0.3μl | 1.50% | 0.15μl | ||
20-200μl | 1μl | 200 | 0.60% | 1.2μl | 0.15% | 0.3μl |
100 | 0.80% | 0.8μl | 0.30% | 0.3μl | ||
20 | 3.00% | 0.6μl | 1.00% | 0.2μl | ||
50-200μl | 1μl | 200 | 0.60% | 1.2μl | 0.15% | 0.3μl |
100 | 0.80% | 0.8μl | 0.30% | 0.3μl | ||
50 | 1.00% | 0.5μl | 0.40% | 0.2μl | ||
100-1000μl | 5μl | 1000 | 0.60% | 6μl | 0.20% | 2μl |
500 | 0.70% | 3.5μl | 0.25% | 1.25μl | ||
100 | 2.00% | 2μl | 0.70% | 0.7μl | ||
200-1000μl | 5μl | 1000 | 0.60% | 6μl | 0.20% | 2μl |
500 | 0.70% | 3.5μl | 0.25% | 1.25μl | ||
200 | 0.90% | 1.8μl | 0.30% | 0.6μl | ||
500-5000μl | 50μl | 5000 | 0.50% | 25μl | 0.15% | 7.5μl |
2500 | 0.60% | 15μl | 0.30% | 7.5μl | ||
1000 | 0.70% | 7μl | 0.30% | 3μl | ||
1-10ml | 0.1ml | 10ml | 0.60% | 60μl | 0.20% | 20μl |
5ml | 1.20% | 60μl | 0.30% | 15μl | ||
2ml | 3.00% | 60μl | 0.60% | 12μl | ||
Note: 1~10ml is not available for micropette plus autoclaveable pipettes.
8-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume(μl) | Maximume Permissible Systematic Error | Maximume Permissible Random Error | ||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15μl | 1.50% | 0.15μl |
5 | 2.50% | 0.125μl | 2.50% | 0.125μl | ||
1 | 4.00% | 0.04μl | 4.00% | 0.04μl | ||
5-50μl | 0.5μl | 50 | 1.00% | 0.5μl | 0.50% | 0.25μl |
25 | 1.50% | 0.375μl | 1.00% | 0.25μl | ||
5 | 3.00% | 0.15μl | 2.00% | 0.1μl | ||
50-300μl | 5μl | 300 | 0.70% | 2.1μl | 0.25% | 0.75μl |
150 | 1.00% | 1.5μl | 0.50% | 0.75μl | ||
50 | 1.50% | 0.75μl | 0.80% | 0.4μl | ||
12-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume(μl) | Maximume Permissible Systematic Error | Maximume Permissible Random Error | ||
0.5-10μl | 0.1μl | 10 | 1.50% | 0.15μl | 1.50% | 0.15μl |
5 | 2.50% | 0.125μl | 2.50% | 0.125μl | ||
1 | 4.00% | 0.04μl | 4.00% | 0.04μl | ||
5-50μl | 0.5μl | 50 | 1.00% | 0.5μl | 0.50% | 0.25μl |
25 | 1.50% | 0.375μl | 1.00% | 0.25μl | ||
5 | 3.00% | 0.15μl | 2.00% | 0.1μl | ||
50-300μl | 5μl | 300 | 0.70% | 2.1μl | 0.25% | 0.75μl |
150 | 1.00% | 1.5μl | 0.50% | 0.75μl | ||
50 | 1.50% | 0.75μl | 0.80% | 0.4μl | ||
Fixed Volume Pipettes | ||||||
Volume Range | Increment | Test Volume(μl) | Maximume Permissible Systematic Error | Maximume Permissible Random Error | ||
5μl | / | 5μl | 1.3% | 0.065μl | 1.2% | 0.06μl |
10μl | / | 10μl | 0.8% | 0.08μl | 0.8% | 0.08μl |
20μl | / | 20μl | 0.6% | 0.12μl | 0.5% | 0.1μl |
25μl | / | 25μl | 0.5% | 0.125μl | 0.3% | 0.075μl |
50μl | / | 50μl | 0.5% | 0.25μl | 0.3% | 0.15μl |
100μl | / | 100μl | 0.5% | 0.5μl | 0.3% | 0.3μl |
200μl | / | 200μl | 0.4% | 0.8μl | 0.2% | 0.4μl |
250μl | / | 250μl | 0.4% | 1.0μl | 0.2% | 0.5μl |
500μl | / | 500μl | 0.3% | 1.5μl | 0.2% | 1.0μl |
1000μl | / | 1000μl | 0.3% | 3.0μl | 0.2% | 2.0μl |
2000μl | / | 2000μl | 0.3% | 6.0μl | 0.15% | 3.0μl |
5000μl | / | 5000μl | 0.3% | 15μl | 0.15% | 7.5μl |